CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm
Limit range
Missing Beat
Multifocal PVCs
Pause
R on T
-12.0 to +12.0
ST XXX high
mm
where XXX = ECG lead
label
-12.0 to +12.0
ST XXX low
mm
where XXX = ECG lead
label
SV Tachy
Tachy
20 to 300 bpm
The Tachy limit
matches the
high heart rate
limit. If the
high heart rate
limit is changed,
the Tachy limit
changes.
Trigeminy
V Brady
V Fib / V Tach
V Tach
VT > 2
Alarm delay specifications for impedance respiration alarms
The following table lists the alarm delays for physiological alarms related to the impedance
respiration measurement. Any alarm delays that are less than 5 s are given as <5 s.
Alarm
Limit range
Apnea
Cardiac Artifact
RR (Imped.) high
4 to 120 breaths
per minute
RR (Imped.) low
4 to 120 breaths
per minute
Alarm delay specifications for SpO
The following table lists the alarm delays for physiological alarms related to the SpO
measurement. Any alarm delays that are less than 5 s are given as <5 s.
2094480-001
Limit increment
Alarm delay
<5 s
8 s
6 s
5 s (average) (5 to 6 s range)
0.1 mm
86 s (average) (82 to 90 s range)
0.1 mm
86 s (average) (85 to 89 s range)
5 s
9 s
1 bpm
10 s
11 s
9 s
7 s
10 s
Limit increment
Alarm delay
3 to 30 s, configurable
30 breaths. The delay time is dependent on
the respiration rate value, for example, with
a respiration rate of 60 breaths per minute,
the delay is 30 s.
< 5 s
1 breath per
minute
< 5 s
1 breath per
minute
alarms
2
CARESCAPE ONE
Alarm specifications
2
265