CARESCAPE ONE User Manual DRAFT 26 October 2017
Default settings
settings can be changed for the current patient and also saved as the channel's profile
setting defaults. For example, if the channel 1 label Art 1 is changed to RAP 1, the RAP
1 label default settings are applied to channel 1 for the current patient. The current
patient settings can be changed and also saved the channel 1 profile settings defaults.
If no label is selected for a channel, then the channel's default settings are used.
Invasive pressure channel 1 defaults to the channel label Art in the ED, ICU, OR, and
PACU software packages. It defaults to the channel label UAC in the NICU software
package. Invasive pressure channel 2 defaults to the channel label CVP in the ED, ICU,
OR, and PACU software packages. It defaults to the channel label UVC in the NICU
software package.
Art 1 pressure site default settings
The default P1 pressure site for the ED, ICU, OR, and PACU software package is Art. Select
Monitor Setup > Main Setup > Parameter Setup > Page 1 > Art 1 > Setup > Page 1 or
Page 2 to adjust these settings for the current patient. The following table lists the factory
default settings.
Setting
Label
Select the pressure site
label.
Scale mmHg
Select the waveform
scale.
Hemo Sweep
Select the waveform
Speed
display sweep speed.
Display Format
Select the pressure
values display format.
Primary HR Source
Select the parameter to
calculate HR.
Show Pulse Rate
Enable the pulse rate
display.
Select Monitor Setup > Main Setup > Parameter Setup > Page 1 > Art 1 > Advanced to adjust
these settings for the current patient. The following table lists the factory default settings.
Setting
Filter Hz
Select the waveform
filter.
Catheter
Enable the disconnect
Disconnect
alarm.
360
ED
Description
(Adolescent profile: 0-170;
Pediatric profile: 0-140;
Infant profile: 0-120)
ED
Description
CARESCAPE ONE
ICU
OR
Art
0-200
0-200
(Perfusion
profile:
0-120;
Adolescent
profile:
0-170;
Pediatric
profile:
0-140; Infant
profile: 0-120
25 mm/s
Sys/Dia (Mean)
AUTO
(US only: ECG)
Disabled
ICU
OR
12
Enabled
PACU
0-200
(Adolescent
profile:
0-170;
Pediatric
profile:
0-140; Infant
profile:
0-120)
PACU
2094480-001