CARESCAPE ONE User Manual DRAFT 26 October 2017
WARNING
WARNING
Batteries must be installed and fully charged prior to taking into use.
Testing the battery charge
Before installing a battery, verify the battery's state of charge. Each battery must be
fully charged before use.
1. Press the TEST button on the battery and check the green charging level indicators
to see how much charge is left:
Four LEDs illuminated: 75% to 100% of full-charge capacity.
Three LEDs illuminated: 50% to 74.9% of full-charge capacity.
Two LEDs illuminated: 25% to 49.9% of full-charge capacity.
One LED illuminated: 11% to 24.9% of full-charge capacity.
One LED flashing: < 11% of full-charge capacity.
Inserting and removing the CARESCAPE ONE battery
WARNING
WARNING
WARNING
CAUTION
2094480-001
PHYSICAL INJURY. Do not install the device above a patient.
Make sure the battery is completely inserted and the battery
door is completely closed. Falling batteries could seriously or
fatally injure neonatal or other vulnerable patients.
EXPLOSION HAZARD. Do not incinerate a battery or store at
high temperatures. Serious injury or death could result.
EXPLOSION OR FIRE. Using non-recommended batteries
could result in injury/burns to the patients or users. Only use
batteries recommended or manufactured by GE. The warranty
can be voided if non-recommended batteries are used.
PHYSICAL INJURY. Do not install the device above a patient.
Make sure the battery is completely inserted and the battery
door is completely closed. Falling batteries could seriously or
fatally injure neonatal or other vulnerable patients.
PHYSICAL INJURY. Do not install the device above a patient.
Leaks from the battery cells can occur under extreme
conditions. The liquid is caustic to the eyes and skin. If the
liquid comes in contact with eyes or skin, flush with clean
water and seek medical attention.
LOSS OF MONITORING. To prevent loss of monitoring, only
change the CARESCAPE ONE battery when CARESCAPE ONE is
connected to a powered CARESCAPE Dock F0.
CARESCAPE ONE
System introduction
43