CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
Alarm
Limit range
FemV X mean low
●
ICP X mean low
●
LAP X mean low
●
P1 mean low to
●
P2 mean low
RAP X mean low
●
RVP X mean low
●
UAC X sys low /
●
UAC X dia low /
UAC X mean low
UVC X mean low
●
-25 to 320 mmHg
CPP X high
(-3.3 to 42.7 kPa)
-25 to 320 mmHg
CPP X low
(-3.3 to 42.7 kPa)
PR(Art) X high
20 to 300 bpm
●
PR(Fem) X high
●
PR(UAC) X high
●
PR(Art) X low
20 to 300 bpm
●
PR(Fem) X low
●
PR(UAC) X low
●
Alarm delay specifications for temperature alarms
The following table lists the alarm delays for physiological alarms related to the temperature
measurement. Any alarm delays that are less than 5 s are given as <5 s.
Alarm
Limit range
T1 high
10 to 45°C (50 to
113°F)
T2 high
T1 low
T2 low
Alarm delay specifications for CO
The following table lists the alarm delays for physiological alarms related to the CO
measurement. Any alarm delays that are less than 5 s are given as <5 s.
Alarm
Limit range
Apnea (CO2)
0 to 13%
EtCO2 high / EtCO2
low
0.0% to 3%
FiCO2 high / FiCO2
low
268
Limit increment
Alarm delay
16 s (average) (14 to 17 s range)
1 mmHg (0.1 kPa)
18 s (average) (16 to 20 s range)
1 mmHg (0.1 kPa)
7 s
1 bpm
7 s
1 bpm
Limit increment
Alarm delay
0.1°C (0.18°F)
61 s (average) (60 to 61 s range)
alarms
2
Limit increment
Alarm delay
20 s (average) (20 s range)
0.1%
21 s (average) (20 to 23 s range)
0.1%
21 s (average) (20 to 23 s range)
CARESCAPE ONE
2
2094480-001