Messages
For information regarding alarm priorities and escalation times, see the supplemental
information provided.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
al. area = alarm area
param. = parameter window
wavef. = waveform area
Message
Analog output
malfunction
Battery empty!
Battery failure
Battery low
Battery temp high
Call Service: Text(s)
missing
Case ended
Case started
Condition battery
Configuration error(s)
DEMO MODE Not for
clinical use!
Entering standby
230
Location
Possible causes
al. area
CARESCAPE ONE analog output
voltage failure.
al. area
The CARESCAPE ONE is being
used on battery power, and
there is less than 5 minutes of
charge left.
al. area
The battery is faulty.
al. area
The CARESCAPE ONE is being
used on battery power, and
there is less than 20 minutes of
charge left.
al. area
Battery temperature is too high.
al. area
The software text is missing in
this language; the text file may
be corrupted.
al. area
OR and PACU software
packages: There is no active
patient case.
al. area
OR and PACU software
packages: A new case has
just been started.
al. area
The battery needs to be
conditioned.
al. area
One or more errors have been
detected in the configuration.
al. area
DEMO mode has been enabled.
al.area
Activate standby has been
selected.
CARESCAPE ONE
Suggested actions
Contact qualified service
personnel.
Charge the battery by
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
Charge the battery by
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
If the problem persists,
contact qualified service
personnel.
Contact qualified service
personnel.
No action required.
No action required.
Replace the battery, and
condition the battery that
was removed.
Contact qualified service
personnel.
To start monitoring a patient,
restart the monitor, or select
Monitor Setup > Defaults
& Service > Exit DEMO >
Confirm.
Continue disconnecting the
patient to enter standby
mode.
2094480-001