CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
For information regarding alarm priorities and escalation times, see the supplemental
information provided.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
al. area = alarm area
●
param. = parameter window
●
wavef. = waveform area
●
Message
Analog output
malfunction
Battery empty!
●
Battery failure
●
Battery low
●
Battery temp high
●
Call Service: Text(s)
●
missing
Case ended
●
Case started
●
Condition battery
●
Configuration error(s)
●
DEMO MODE Not for
●
clinical use!
Entering standby
●
230
Location
Possible causes
al. area
CARESCAPE ONE analog output
●
voltage failure.
al. area
The CARESCAPE ONE is being
●
used on battery power, and
there is less than 5 minutes of
charge left.
al. area
The battery is faulty.
●
al. area
The CARESCAPE ONE is being
●
used on battery power, and
there is less than 20 minutes of
charge left.
al. area
Battery temperature is too high.
●
al. area
The software text is missing in
●
this language; the text file may
be corrupted.
al. area
OR and PACU software
●
packages: There is no active
patient case.
al. area
OR and PACU software
●
packages: A new case has
just been started.
al. area
The battery needs to be
●
conditioned.
al. area
One or more errors have been
●
detected in the configuration.
al. area
DEMO mode has been enabled.
●
al.area
Activate standby has been
●
selected.
CARESCAPE ONE
Suggested actions
Contact qualified service
●
personnel.
Charge the battery by
●
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
●
Charge the battery by
●
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
●
If the problem persists,
●
contact qualified service
personnel.
Contact qualified service
●
personnel.
No action required.
●
No action required.
●
Replace the battery, and
●
condition the battery that
was removed.
Contact qualified service
●
personnel.
To start monitoring a patient,
●
restart the monitor, or select
Monitor Setup > Defaults
& Service > Exit DEMO >
Confirm.
Continue disconnecting the
●
patient to enter standby
mode.
2094480-001