CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
Multifocal PVCs
Noisy ECG
Noise
Pause
R on T
ST XXX high / ST XXX
low
where XXX = ECG lead
label
SV Tachy
Trigeminy
V Brady
V Fib / V Tach
V Tach
VT > 2
212
Location
Possible causes
al. area,
Physiological alarm.
wavef.
al. area
Arrhythmia alarm category
has been set to Off through
wavef.
Alarms Setup > Arrhythmia >
Setup, and the ECG channels
have not been available for
analysis for the last 20 seconds
due to excessive noise; the
internal HR calculation has not
been updated for the last 30
seconds, which compromises
the accuracy of detecting
events; or in single lead analysis
when arrhythmia detection is
set to Off, the ECG 1 lead has
not been active for the last 5
seconds.
al. area,
Physiological alarm.
wavef.
al. area,
Physiological alarm.
wavef.
al. area
Measurement values are equal
to or outside the alarm limits.
al. area,
Physiological alarm.
wavef.
al. area,
Physiological alarm.
wavef.
al. area,
Physiological alarm.
wavef.
al. area,
Physiological alarm.
param.,
wavef.
al. area,
Physiological alarm.
param.,
wavef.
al. area,
Physiological alarm.
wavef.
CARESCAPE ONE
Suggested actions
Check the patient status.
Check and remove the source
of excessive noise if possible.
Check the patient status.
Check the patient status.
Check the patient status.
Adjust alarm limits if
necessary.
Check the patient status.
Check the patient status.
Check the patient status.
Check the patient status.
Check the patient status.
Check the patient status.
2094480-001