CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
Artifact
●
Check device
●
SpO2 malfunction
●
Check probe
●
Check SpO2 probe
●
Demo mode
●
SpO2 Demo mode
●
Expiring cable
●
Expiring probe
●
Faulty cable
●
SpO2 faulty cable
●
Faulty probe
●
SpO2 faulty probe
●
Faulty adhesive SpO2
●
probe
Identical SpO2 device
●
Incompatible cable
●
Incompatible SpO2
●
cable
2094480-001
Location
Possible causes
param.
●
Artifact detected.
param.
●
CARESCAPE SpO
malfunction. No SpO
al. area
●
detected.
param.
●
There is no detectable SpO
signal, the sensor is faulty or is
al. area
●
detached from the patient.
With CARESCAPE SpO
this message also appears
when the saturation updates
exceed 30 seconds, the pulse
rate is lost, or the pulse timeout
condition occurs.
param.
●
CARESCAPE SpO
demo mode.
al. area
●
param.
●
CARESCAPE SpO
detected that the interconnect
cable is near its expiration date.
param.
●
CARESCAPE SpO
has detected that the sensor
or adhesive sensor is near its
expiration date.
param.
●
CARESCAPE SpO
has detected a defective
al. area
●
interconnect cable.
param.
●
The sensor has failed.
The adhesive sensor has failed.
al. area
●
al. area
You can only have one
●
CARESCAPE SpO
SpO
– Masimo, or CARESCAPE
2
SpO
– Nellcor in the system.
2
param.
●
CARESCAPE SpO
detected that an incompatible
al. area
●
or unrecognized interconnect
cable has been connected.
CARESCAPE ONE
Suggested actions
Check sensor contact.
●
Reposition or replace sensor.
●
Request the patient to remain
●
still.
device
Check the CARESCAPE SpO
●
2
data
device.
2
Replace if needed.
●
Check the sensor and
●
2
connections.
Check the patient status.
●
– Nellcor
2
– Masimo is in
Exit Demo mode. See the
●
2
supplemental information
provided.
– Masimo has
Replace the interconnect
●
2
cable.
– Masimo
Replace the sensor or
●
2
adhesive sensor.
– Masimo
Replace the interconnect
●
2
cable.
Replace the sensor.
●
Replace the adhesive sensor.
●
Remove all but one
●
, CARESCAPE
CARESCAPE SpO
2
– Masimo has
Connect a Masimo
●
2
compatible interconnect
cable.
Messages
2
device.
2
215