CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
MESSAGE
NOTES
LA/L lead off
LL/F lead off
RA/R lead off
RL/N lead off
V/C lead off
Va/Ca lead off
Vb/Cb lead off
V2/C2 lead off
V3/C3 lead off
V4/C4 lead off
V5/C5 lead off
V6/C6 lead off
Lead changed
Lead off
Leads off
Alarm priority =
Escalating
Leads off
Learning
Missing Beat
Move ECG device
to a cooler
location
Multifocal PVCs
Noisy ECG
Noise
Noisy ECG
Alarm priority =
Escalating
Noise
Pause
R on T
ST XXX high / ST
XXX low where
XXX = ECG lead
label
276
LOCATION
PRIORITIES AND ESCALATION TIMES
info
0 s
wavef.
wavef.
7 s
alarm
alarm,
wavef.
alarm,
wavef.
0 s
wavef.
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm
wavef.
alarm
wavef.
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm
60s, according to priority setting: high, medium, low,
info
CARESCAPE ONE
low
medium
0 s
7 s
3 s
12 s (30
s in OR
software
package)
0s, according to priority setting: high,
medium, low
0 s
0 s, according to priority setting: high,
medium, low
5 s
12 s
high
60 s
2094480-001