CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
No probe
●
No SpO2 probe
●
No adhesive SpO2
●
probe
PR(SpO2) high /
●
PR(SpO2) low
Probe initializing
●
Probe off
●
SpO2 Probe Off
●
Pulse search
●
Replace cable
●
Replace SpO2 cable
●
Replace probe
●
Replace SpO2 probe
●
Replace adhesive
●
SpO2 probe
SpO2 device failure:
●
Call service
SpO2 device
●
overheated Shutting
down
2094480-001
Location
Possible causes
param.
●
Sensor is not connected to the
CARESCAPE SpO
al. area
●
Adhesive sensor is not
connected to the CARESCAPE
SpO
device.
2
al. area
Measurement values are equal
●
to or outside the alarm limits.
param.
●
CARESCAPE SpO
has detected that the sensor is
initializing.
param.
●
The sensor may be defective or it
is not connected to the patient.
al. area
●
param.
●
The measurement is starting
normally.
Defective or damaged sensor or
cable.
Sensor is off of the patient.
Detection of a repeatable pulse
has stopped.
Pulse is not recognized by
the algorithm due to a low
amplitude signal.
param.
●
CARESCAPE SpO
detected that the interconnect
al. area
●
cable has expired.
param.
●
CARESCAPE SpO
detected that the sensor has
al. area
●
expired.
CARESCAPE SpO
has detected that the adhesive
sensor has expired.
al. area
CARESCAPE SpO
●
or initialization failure.
al. area
The surface temperature of the
●
CARESCAPE SpO
the CARESCAPE ONE turns the
CARESCAPE SpO
CARESCAPE ONE
Suggested actions
Check the connection
●
device.
between the sensor and
2
the CARESCAPE SpO
Check the patient status.
●
Adjust alarm limits if
●
necessary.
– Masimo
SpO
●
2
2
after initialization is complete.
Check the patient status.
●
Reposition the SpO
●
Replace the SpO
●
Wait for waveforms and
●
parameter values to display.
Check the sensor and cable.
●
Reposition or replace sensor.
●
– Masimo has
Replace the interconnect
●
2
cable.
– Masimo has
Replace the sensor.
●
2
– Masimo
2
device startup
Reconnect the CARESCAPE
●
2
SpO
2
Replace the CARESCAPE SpO
●
If the problem persists,
●
contact qualified service
personnel.
Check the CARESCAPE SpO
●
is too high and
device placement and ensure
2
that it is not covered by
off.
anything that might lead to
2
overheating.
Replace the CARESCAPE SpO
●
device.
Messages
device.
2
monitoring will begin
sensor.
2
sensor.
2
.
.
2
2
2
217