The effects of interfering gases and vapors shall be as follows:
ISO 806061-2-55 Table 201.105 Test gas levels of interfering gases and vapors in nominal atmospheric
pressure, i.e., 760 mmHg:
Effects of interfering gases and vapors
Effects of non-interfering gases and
vapors
322
Effect of Helium (50% vol): decreases CO
Effect of N
O (30% vol): increases CO
2
Effect of O
(40% to 95%):
2
If compensation is not activated: decreases CO
vol%
If compensation is activated: CO
Effect of N
O (30% to 80%):
2
If compensation is not activated: CO
If compensation is activated: CO
A gas is non-interfering if its effect to the measured CO
follows:
0-40 mmHg: ±1 mmHg additional error
41-70 mmHg: ±2.5% additional error
71-100 mmHg: ±4% additional error
101-150 mmHg: ±5% additional error
The following gases are non-interfering when tested per ISO