Hpe5022_Ber_Prbspattern - Agilent Technologies E5022B Programming Manual

Hard disk read/write test system
Hide thumbs Also See for E5022B:
Table of Contents

Advertisement

Function Reference
BER Data Pattern Definition Function

hpe5022_BER_prbsPattern

C Syntax
ViStatus hpe5022_BER_prbsPattern(ViSession id, ViInt16 pat, ViInt32 rand_coef,
ViInt32 rand_init);
Visual Basic
hpe5022_BER_prbsPattern(ByVal id As Long, ByVal pat As Integer, ByVal
Syntax
rand_coef As Long, ByVal rand_init As Long) As Long
Description
This function configures the PRBS (pseudo random bit sequence) pattern and
assigns the data pattern number. The "hpe5022_BER_selectPattern" function
selects the data pattern number to be used for bit error measurement.
Parameters
• id
• pat
• rand_coef
• rand_init
34
Description
Specifies the system identifier. This is given by the
"hpe5022_init" function.
Direction
IN
Description
Selects the data pattern number to be assigned.
Direction
IN
Values
Same as 'pat' parameter in the
"hpe5022_BER_selectPattern" function.
Description
Specifies the coefficients of the pseudo random sequence's
generating polynomial. LSB represents the 0'th order
coefficient of the generating polynomial. This function can
generate a pseudo random data up to a maximum of 16th
order.
Direction
IN
Preset Value
137 (X
Values
Name
hpe5022_BER_RAND_COEF_MIN
hpe5022_BER_RAND_COEF_MAX
Description
Specifies the initial value of the pseudo random generator.
Direction
IN
7
3
+X
+1)
Value
3
65535