Monitoring The Clock And Environmental Sensors - A&D BM series Instruction Manual

Analytical balance
Hide thumbs Also See for BM series:
Table of Contents

Advertisement

Monitoring the Clock and Environmental Sensors

6.6.
In the weighing mode, the built-in clock and environmental sensors (temperature, humidity,
barometric pressure) can be monitored and temperature data can be output in the function
settings.
Specifications of the environmental sensors
Sensor
Temperature
Humidity
Barometric pressure
Notes
The environmental sensors are built into the balance casing. Therefore, these values are
not in accordance with the environmental condition.
Refer to "20. Specifications" for the balance operating conditions.
Operation
Press the
SELECT
mode. The current time, date and sensor
values (temperature, humidity, barometric
pressure) are displayed for several seconds
in order. The balance returns to the
weighing mode automatically.
When pressing the
while the current data is displayed,
the display proceeds to the next
data.
When pressing the
while the current data is displayed,
the balance returns to the weighing
mode.
Notes
When adjusting the built-in clock, refer to "10.7. Clock and Calendar Function".
When adding the time and date to the output data, refer to "Time/Date output (5-td)" of
"10.2. Details of the Function Table".
When adding temperature data to the output data, refer to "Temperature output (5-tp)"
of "10.2. Details of the Function Table".
Resolution
Measurement range
±1.5 °C
5 °C to
±10 %
0 %
±10 hPa
300 hPa to 1100 hPa
key in the weighing
key
SELECT
key
CAL
Applicable range
40 °C
5 °C to
to 100 %
Weighing mode
Time
5 seconds
Date
5 seconds
Humidity
Environmental sensors
10 seconds
Weighing mode
32
40 °C
Barometric pressure
Temperature
BM series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bm-22Bm-20Bm-300Bm-500Bm-252Bm-200