Initializing The Balance - A&D BM series Instruction Manual

Analytical balance
Hide thumbs Also See for BM series:
Table of Contents

Advertisement

Initializing the Balance

9.2.
This function restores the following parameters to the factory settings.
Calibration data
Function table (excluding the time and date)
Sample unit mass value (counting mode),
100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External weight value
Function switch settings ("9.1. Permit or Inhibit")
Liquid density and temperature in the density mode
Note Be sure to calibrate the balance after initialization.
1 Press the
ON:OFF
2 While pressing and holding the
key, press the
RANGE
3 Press the
RANGE
4 Press the
PRINT
To cancel the operation, press the
5 Press the
RE-ZERO
6 Press the
PRINT
The balance will automatically return to the weighing mode.
key to turn the display off.
PRINT
key to display
ON:OFF
key to display
Clr
key to display
Clr no
CAL
key to display
Clr go
key to initialize the balance.
key and the
.
p5
.
.
key.
.
44
Press and hold Press
BM series