15. Password Function

The password function can be used to restrict balance use or functions. It is useful for preventing
the time and date setting from being falsified and preventing users from changing other function
settings.
Password: Four digits (4 x 4 x 4 x 4 = 256 outcomes) using the
RE-ZERO
At the factory setting, the password function has been disabled.
Settings related to the password function are performed in the function table.
Note The password function varies depending on the balance software version.

Balance Software Version 1.50 to 1.64

15.1.
Two settings are available in "Password function (LocFnc) " of the function table.
No password function.
pA55 0
Entering a password is required before starting weighing.
pA55 1
pA55 0 (No password function)
No password is required. Anybody can perform weighing, use all the balance functions and make
changes to them.
pA55 1 (Entering a password is required before starting weighing.)
The administrator (ADMIN ) sets a unique password to restrict balance users. Only one
password by the administrator can be set.
The correct password must be entered for the balance to enter the weighing mode.