Calibration Using the Internal Mass

8.3.
This function calibrates the balance using the internal mass.
The only operation required is to press the
1 Connect the AC adapter and warm up the balance for at
least one hour with nothing on the weighing pan.
2 Press the
CAL
3 The balance performs calibration using the internal mass.
Do not allow vibration or drafts to affect the balance.
4
GLP output is set,
If
output to the RS-232C interface and is stored in memory. Refer to
"GLP output (
info
table in "11.2. GLP Report" and "12. Data Memory".
is displayed after calibration.
end
5 The balance will automatically return to the weighing mode after calibration.
6 Confirm weighing accuracy using calibration test (
key to display
Cal in
is displayed, the calibration report is
glp
)" and "Data memory (
key.
CAL
.
)" of the function
data
).
CC in
38
GLP output
BM series