Calibration report using an external weight
Key operation
1 Press and hold the
2
is displayed.
Cal 0
3 To update the calibration weight value, press the
To use the preset calibration weight value, proceed to Step 5.
4 Specify the calibration weight value using the following keys.
key ........ To select the blinking digit.
RANGE
(+)key ... To increase the value of the blinking digit.
RE-ZERO
(-)key ....... To decrease the value of the blinking digit.
MODE
key ......... To store the new value.
PRINT
5 Press the
PRINT
seconds.
6 Place the displayed weight on the pan. Press the
weight value is displayed for a few seconds.
7
is displayed. Remove the weight from the pan.
end
8 If GLP output is used,
9 The balance returns to the weighing mode.
Function setting "info 1"
AD-8121 format
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
----+200.0000--g
SIGNATURE
---------------
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
key (for approx. 2 seconds) to display
CAL
key. The zero point is measured and the weight value is displayed for a few
is displayed and the calibration report is output.
glp
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
Cal out
key and proceed to Step 4.
RANGE
key to measure the weight and the
PRINT
Function setting "info 2"
General format using the balance built-in clock data