Calibration test report using an external weight
Note Calibration test does not perform adjustment.
Key operation
1 Press and hold the
2
is displayed.
Cal 0
3 To update the target weight value, press the
To use the preset target weight value, proceed to Step 5.
4 Specify the target weight value using the following keys.
key .......... To select the blinking digit.
RANGE
(+)key .... To increase the value of the blinking digit.
RE-ZERO
(-)key ......... To decrease the value of the blinking digit.
MODE
key ............ To store the new value.
PRINT
5 Press the
PRINT
6 Place the displayed target weight on the pan. Press the
value is displayed for a few seconds.
7
is displayed. Remove the weight from the pan.
end
8
is displayed and the calibration test report is output.
glp
9 The balance returns to the weighing mode.
Function setting "info 1"
AD-8121 format
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CAL.TEST(EXT.)
ACTUAL
-------0.0000--g
----+200.0002--g
TARGET
----+200.0000--g
SIGNATURE
---------------
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
BM series
key (for approx. 2 seconds) to display
CAL
key. The zero point is measured and the weight value is displayed for a few seconds.
Manufacturer
Serial number
ID number
Calibration test type
Zero point value
Target weight value
Target weight
Signature
key and proceed to Step 4.
RANGE
PRINT
Model
Date
Time
63
and release the key.