Correcting The Internal Mass Value - A&D BM series Instruction Manual

Analytical balance
Hide thumbs Also See for BM series:
Table of Contents

Advertisement

Correcting the Internal Mass Value

8.7.
The balance can correct the internal mass
value within the range shown below. This
function corrects the internal mass value to
conform to an external weight. The corrected
mass value is maintained in non-volatile
memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
BM-20, BM-22
BM-252, BM-500, BM-300, BM-200
Example: With the BM-300, 200.0000 g is corrected by +0.6 mg (200.0006 g). When performing a
+0.6 mg correction with a 100 g external weight, the correction value will be +1.2 mg in
accordance with the reference value of 200 g.
1 Perform calibration using the internal mass. Weigh an external
weight and get the correction value.
2 Press the
ON:OFF
3 While pressing and holding the
press the
ON:OFF
4 Press the
PRINT
the function table switch and internal mass correction switch
to "
" as shown to the right using the following keys.
1
key .......... To select the blinking digit.
RANGE
key ....... To change the value of the blinking digit.
RE-ZERO
key ............ To store the new value and return to the weighing mode.
PRINT
key ............... To cancel the current operation.
CAL
5 Press and hold the
the function table and release the key when
6 Press the
RANGE
7 Press the
PRINT
Correct the internal mass value using the following keys.
(+)key .... To increase the value.
RE-ZERO
(-)key ........ To decrease the value.
MODE
key ........... To store the new value and display the next
PRINT
key ............ To cancel the current operation and display
CAL
8 Press the
CAL
9 Press the
CAL
internal mass.
10 Check that the correction has been performed properly with the
external weight. If the value is incorrect, repeat the correction.
Model
key to turn the display off.
PRINT
key.
is displayed.
p5
key to display the function switches. Set
Internal mass correction switch
key (for approx. 2 seconds) to enter
RANGE
key several times to display
key.
menu item of the function table.
the next menu item of the function table.
key to return the weighing mode.
key to calibrate the balance using the
200 g
200. 0 000
g
Correction reference value Correction range
20.00000 g
200.0000 g
key and the
RANGE
Function table switch
is displayed.
ba5fnc
.
C5 in
42
The same mass
Correct the internal mass
by +0.6 mg at 200 g.
Calibrate with this
internal mass.
Corrected value
±1.5 mg
±1.5 mg
Standby
Press and hold Press
key,
Function switches
Weighing
mode
Press and hold
(approx. 2 seconds)
Function table
Internal mass correction
Corrected value
Weighing mode
Perform calibration
using the internal mass
200 g
200. 0 006
g
BM series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bm-22Bm-20Bm-300Bm-500Bm-252Bm-200

Table of Contents