A&D BM series Instruction Manual page 63

Analytical balance
Hide thumbs Also See for BM series:
Table of Contents

Advertisement

Calibration test report using the internal mass
Note Calibration test does not perform adjustment.
Key operation
1 Press and hold the
2
is displayed and the balance is checked automatically.
CC
3 The zero point is measured and the weight value is displayed for a few seconds.
4 The internal mass is weighed and the weight value is displayed for a few seconds.
5
is displayed and the calibration test report is output.
glp
6 The balance returns to the weighing mode.
Command
This calibration test can be performed with the TST command.
Function setting "info 1"
AD-8121 format
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CAL.TEST(INT.)
ACTUAL
-------0.0000--g
----+200.0002--g
TARGET
----+200.0000--g
SIGNATURE
---------------
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
BM series
key (for approx. 2 seconds) to display
CAL
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
CC in
Function setting "info 2"
General format using the balance built-in clock data
61
and release the key.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bm-22Bm-20Bm-300Bm-500Bm-252Bm-200

Table of Contents