Data Memory For Calibration And Calibration Test Data - A&D BM series Instruction Manual

Analytical balance
Hide thumbs Also See for BM series:
Table of Contents

Advertisement

Deleting all memory data at one time
1 Press and hold the
until
is displayed, and then release the key.
reCall
2 Press the
RANGE
.
Clear
3 Press the
PRINT
blinking.
4 Press the
RE-ZERO
"go" blinking.
5 Press the
PRINT
5 The balance displays
weighing mode.

Data Memory for Calibration and Calibration Test Data

12.3.
Features
Calibration data (when and how it is performed) and calibration test
data can be stored in memory.
All the data in memory is available to be output at one time to a printer
or personal computer.
Up to 50 data sets of the latest calibration or calibration test can be
stored. When the memory capacity has been reached, "
displayed in the upper left of the display.
Storing the calibration and calibration test data
Note If "
" appears blinking in the upper left of the display, delete the data in memory.
Clr
Store the calibration and calibration test data as follows:
1 Set the "Data memory (
2 Set the "GLP output (
3 With the settings above, each time calibration or calibration test is performed, the
data is stored automatically.
BM series
key (for approx. 2 seconds)
PRINT
key several times to display
key to display
Clr " n " o
key to display
key to delete all data
and returns to the
end
)" parameter to "
data
)" parameter to "
info
with "no"
with
Clr " g " o
fUl
" (to store the calibration data).
2
", "
" or "
".
1
2
3
69
Press and hold
(approx. 2 seconds)
Press several times
" is
Indicator