Electronic Fuse of SSV Open

Error ID: 2049
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the module has drawn excessive current
causing the selection-valve electronic fuse to open.
Probable cause
1
2
3

AIV Fuse

Error ID: 2044
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the module has drawn excessive current
causing the inlet-valve electronic fuse to open.
Probable cause
1
2
3
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Defective solvent selection valve.
Defective connection cable (front panel to
main board).
Defective main board.
Defective active inlet valve.
Defective connection cable (front panel to
main board).
Defective main board.
Error Information
Module Error Messages
Suggested actions
Restart the pump. If the error message appears
again, exchange the solvent selection valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Restart the module. If the error message
appears again, exchange the active inlet valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.