Test Functions and Calibration
Binary Pump Solvent Compressibility Calibration

Running the Solvent Compressibility Calibration

Running the Solvent Compressibility Calibration from the Agilent Lab
Advisor
When
If a solvent is not available in the list of pre-calibrated solvents, the solvent compressibility
calibration allows to generate appropriate compressibility data.
Tools required
p/n
8710-0510
Parts required
p/n
G1312-67500
Preparations
Place a bottles with solvent to be calibrated in channel A (resp. A1 if a solvent selection valve is
installed).
Avoid inaccurate pump elasticity calibration.
C AU T I O N
This would lead into invalid and not-portable solvent compressibility data.
Make absolutely sure that the pump is very thoroughly flushed with the solvent to be
N O T E
calibrated before starting the procedure! Any trace of other solvents or the smallest air
bubble inside the flow path definitely will cause the calibration to fail!
1 Select the solvent from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
N O T E
completed. Otherwise the pump may generate an overpressure error.
104
Description
Wrench, 1/4 - 5/16 inch, open
Description
Calibration capillary assembly
Make sure to perform accurate pump elasticity calibration.