3
Remove the seal cap from the purge valve.
5
Place a new frit into the purge valve with the orientation
of the frit as shown below (slit in frit points to the front).
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Exchange the Purge Valve Frit or the Purge Valve
4
Using a pair of tweezers or a toothpick remove the frit.
6
Reinstall the seal cap including the gold seal.
N O T E
Before reinstallation always check the gold seal in
the seal cap. A deformed seal cap should be
exchanged.