Exchange the Seal Wash Cartridge

Exchange the Seal Wash Cartridge

Parts required
p/n
5065-4445
Preparations
• Switch off pump at the main power switch.
• Open the doors.
1
Remove the wash solvent tubings from the support ring
outlet and from the adapter leading to the waste bottle.
140
Description
Peristaltic pump with Pharmed tubing
2
Unclip the peristaltic pump cartridge from the module
housing and remove it.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual