Usb (Universal Serial Bus) - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

11
Hardware Information
Interfaces
5V Distribution (Future Use)
• Available directly after turn on oft the hosting module (assures that certain
• For digital circuits or similar.
• Provided 500 mA maximum.
• Short-circuit proof with automatic switch off (by firmware).
24V Distribution (Future Use)
• Available by firmware command (defined turn on/off).
• For devices that need higher power
• Class depends on hosting module's internal power overhead.
• If a connected device requires more power the firmware detects this
• Fuse used for safety protection (on board).
• Short circuit will be detected through hardware.

USB (Universal Serial Bus)

USB (Universal Serial Bus) - replaces RS232, supports:
• a PC with control software (for example Agilent Lab Advisor)
• USB Flash Disk
196
base functionality of the device can be detected by firmware).
Class 0: 0.5 A maximum (12 W)
Class 1: 1.0 A maximum (24 W)
Class 2: 2.0 A maximum (48 W)
(overcurrent detection) and provides the information to the user interface.