Using the plastic side of the insert tool, insert new seals
into the pump head.
9
Place the pump housing onto the piston housing.
11
Tighten the lock screw.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Maintenance of a Pump Head without Seal Wash
8
Place the support rings on the piston housing. Note the
correct position of the pins.
10
Insert the pistons and carefully press them into the
seals.
12
Install the pump head, see