Maintenance
Exchange the Purge Valve Frit or the Purge Valve
7
Insert the purge valve into the purge valve holder.
122
C AU T I O N
Damage to the purge valve
Do not lift the pump using the purge valve as a handle,
it might get leaky.
Do not try to turn the purge valve into the correct
position when already fixed to the pump. The rubber
o-ring might break.
Anticipate the correct position of the connections
before tightening the valve.
8
Tighten the purge valve and reconnect outlet capillary
and waste tubing.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual