Agilent Technologies 1260 Infinity II User Manual page 31

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Table 2
Type
Composition range
Composition
precision
Composition
accuracy
Integrated degassing
unit
Instrument Control
Communications
Safety and
maintenance
GLP features
Housing
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Performance Specifications 1260 Infinity II Binary Pump (G7112B)
Specification
settable: 0 – 100 %
recommended: 1 – 99 % or 5 µL/min per
channel, whichever is greater
< 0.15 % RSD or < 0.04 min SD,
whichever is greater
± 0.35 % absolute
Number of channels: 2
Internal volume per channel: 1.5 mL
Agilent control software with LC and CE
Drivers A.02.14 or above
Lab Advisor B.02.09 or above
Agilent Instant Pilot (G4208A) with
firmware B.02.20 or above
Instrument Control Framework (ICF)
A.02.04 or above
Controller-area network (CAN), Extended
Remote Interface (ERI), Local Area
Network (LAN)
Extensive diagnostics, error detection
and display through Agilent LabAdvisor,
leak detection, safe leak handling, leak
output signal for shutdown of the
pumping system. Low voltage in major
maintenance areas.
Early maintenance feedback (EMF) for
continuous tracking of instrument usage
in terms of seal wear and volume of
pumped mobile phase with pre-defined
and user settable limits and feedback
messages. Electronic records of
maintenance and errors.
All materials are recyclable
Site Requirements and Specifications
Performance Specifications
Comments
at 0.2 and 1 mL/min; based
on retention time at constant
temperature
at 2 mL/min, at 10 MPa
(100 bar, 1450 psi)
(water/water with tracer)
For details about supported
software versions refer to the
compatibility matrix of your
version of the LC and CE
Drivers