Manual Configuration; With Telnet - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

12
LAN Configuration

Manual Configuration

Manual Configuration

Manual configuration only alters the set of parameters stored in the
non-volatile memory of the module. It never affects the currently active
parameters. Therefore, manual configuration can be done at any time. A
power cycle is mandatory to make the stored parameters become the active
parameters, given that the initialization mode selection switches are allowing
it.
Figure 30

With Telnet

Whenever a TCP/IP connection to the module is possible (TCP/IP parameters
set by any method), the parameters may be altered by opening a Telnet
session.
1 Open the system (DOS) prompt window by clicking on Windows START
212
Manual Configuration (Principle)
button and select "Run...". Type "cmd" and press OK.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual