Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

3
Carefully plug in the connector of the solvent selection
valve into the internal connection socket on the left side
of the slot behind the pump front metal panel.
5
Connect the outlet of solvent A (upper row) to the top
degasser inlet and the outlet of solvent B (lower row) to
the bottom degasser inlet. Put the solvent bottles into
the solvent cabinet. Connect the bottle heads of solvents
A1 and A2 to the inlets in the upper row, see labels on
valve assembly. Connect the bottle heads of solvents B1
and B2 to the inlets in the lower row, see labels on valve
assembly.
Block unused channels of the SSV using a plug (Blank plug (5041-8365)) to avoid leaks or
N O T E
air entering the solvent channels.
Before using of the system with a new installed valve it may be required to pump several
N O T E
mL of solvent to get the flow stabilized at a pressure ripple as low as it used to be when
the system was still working properly.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Installation of the Solvent Selection Valve Upgrade Kit
4
Install the solvent selection assembly by fixing the
indicated screws with distance sheet under the SSV
body.
8
Maintenance
147