8
Maintenance

Overview of Maintenance and Simple Repairs

Overview of Maintenance and Simple Repairs

Figure 10
pump. The pump heads and its parts require normal maintenance (for
example, seal exchange) and can be accessed from the front (simple repairs).
Replacement of valve cartridges or filters don't require to remove the pump
from the system stack.
Figure 10
1
2
3
4
5
116
on page 116 shows the main user accessible assemblies of the binary
Overview of Maintenance and Simple Repairs
Purge valve, see
"Exchange the Purge Valve Frit or the Purge Valve"
Outlet valve, see
"Exchange the Outlet Valve"
Active inlet valve, see
"Exchange the Active Inlet Valve (AIV) or its Cartridge"
page 137
Pump head, see
"Maintenance of a Pump Head with Seal Wash"
Solvent selection valve, see
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
on page 144
"Exchange the Solvent Selection Valve"
on page 120
on
on page 130
on page 148