Maintenance of a Pump Head without Seal Wash

When
In case of maintenance or pump head internal leaks
Tools required
p/n
8710-0510
01018-23702
Parts required
#
1
OR
1
1
Preparations
• Switch off the pump at the main power switch.
• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters for avoiding leakages.
• Remove the Pump Head Assembly.
Both pump head assemblies use the same internal components. In addition, pump head A
N O T E
is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Description
Open-end wrench 1/4 — 5/16 inch
Hexagonal key, 4 mm
Insert tool
p/n
Description
5063-6589
Piston seal PTFE, carbon filled, black (pack of 2), default
0905-1420
PE seals (pack of 2)
5063-6586
Sapphire piston

Maintenance of a Pump Head without Seal Wash

8
Maintenance
127