Figure 2
Damper and mixer can be bypassed for lowest delay volume of the binary
pump. This configuration is recommended for low flow rate applications with
steep gradients.
Figure 3
instructions on how to change between the two configurations, see
the Binary Pump to Low Delay Volume Mode"
Bypassing the mixer while the damper remains in line is not a supported configuration and
N O T E
may lead to undesired behavior of the binary pump.
12
The hydraulic path of the Binary Pump with damper and mixer
on page 13 illustrates the flow path in low delay volume mode. For
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
on page 64.
"Convert