Initialization Failed - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Initialization Failed

Error ID: 2207, 2217
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable cause
1
2
3
Degasser: signal fail
Error ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser:
• No valid pressure signal is shown during startup of the degasser
• Measured pressure is higher than 32000 hPa
Probable cause
1
2
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Blocked (passive or active) inlet valve.
Defective pump drive assembly.
Defective main board.
Degasser sensor defect
Degasser sensor not connected to main
board.
Error Information
Module Error Messages
Suggested actions
Exchange the (passive or active) inlet valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
6