Leak Sensor Open

Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable cause
1
2
3
4

Leak Sensor Short

Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable cause
1
2
3
4
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Leak sensor not connected to the Power
Switch board.
Defective leak sensor.
Leak sensor incorrectly routed, being pinched
by a metal component.
Power switch assembly defective
Defective leak sensor.
Leak sensor incorrectly routed, being pinched
by a metal component.
Power switch assembly defective
Cable or contact problem.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Suggested actions
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Please contact your Agilent service representative.