Formulation - Sartorius Factory Series FC06BBE-S Operating Instructions Manual

Industrial weighing tech
Hide thumbs Also See for Factory Series FC06BBE-S:
Table of Contents

Advertisement

Formulation

M
Purpose
With this application program you
can add weight values and
calculated values to a totalization
memory as components of a formula.
You can use this application program
in combination with a program
chosen from Application 1 (e.g.,
counting, weighing in percent) and
one from Application 2 (check-
weighing, time-controlled functions).
Available Features
– Totalization of weight values and
calculated values
– Simultaneous storage of net and
calculated values
– Optional configuration in the
Setup menu for loading weight
values and calculated values
either from Application 1 (e.g.,
counting, weighing in percent)
or from Application 2 (check-
weighing, time-controlled
functions)
– Totalization memory for up to
65535 values
– Transaction counter and current
total displayed in the text line
– Scale tared after a value
is stored
– Manual input of the number of
individual weighing operations
(enter on the numeric and press
). Result printed and
nDef
memory cleared when number
entered has been reached
– Optional configuration in the
Setup menu to add the current
weight, with display accuracy,
to the current total by pressing
the
soft key and generating
M+
a printout of the result
– Optional configuration in the
Setup menu for stability-depen-
dent storage of the measured
value:
Stability range
72
– Optional automatic storage of
measured values
Storage of measured value is
indicated by
.
Uu
indicates that you can place
zz
a load on the scale.
– Minimum load threshold for
automatic storage
– Press the
soft key to delete
M-
the last value added to the
totalization memory.
The transaction counter value
is reduced by one and a printout
is generated.
– Printout of an evaluation of
results, depending on the
Application 1 or Application 2
parameters. Configure the Setup
menu to define the information
included on this printout.
– Press
for a printout of an
MR
intermediate evaluation after each
addition or a final evaluation
– Optional configuration in the
Setup menu to clear the
totalization memory and reset
the transaction counter by
pressing c or after an evalua-
tion is printed out
– Totalization data and transaction
counter data are stored in non-
volatile memory
– Continue formulation after turning
the scale off and back on
Factory Settings
Automatic storage:
Off
Minimum load for automatic storage:
10 digits
Source of data for auto storage:
Application 1
Evaluated values:
Net
Evaluation mode, MR key function:
Intermediate
evaluation, print
Printout of individual components:
On
Stability range:
1-3-4 2 increments
Tare after individual printout:
Soft Key Functions
Add weight values or
M+
application values to the
total in the totalization
memory. The component
or transaction counter
value increases by one
each time you press
this key.
Delete the last value
M-
added to memory. The
transaction counter value
decreases by 1. You
cannot delete previous
values by repeatedly
pressing this key.
Print or display an
MR
intermediate or final
evaluation
Store the input number
nDef
of components
No

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents