isoCAL:
Automatic Calibration
and Adjustment
First set either
On and reset
or
the application
without resetting the
(factory setting) in the Setup:
app.
Scale menu.
The "isoCAL" display automatically
begins flashing if the ambient
temperature changes in relation to
the temperature at the time of the last
calibration/adjustment, or after
a defined time interval has elapsed.
The scale is telling you that it wants to
adjust itself.
This automatic calibration prompt
is triggered when:
– The change in temperature is
greater than 10 Kelvin
– The scale status does
not correspond to Setup
configurations
– No number or letter input
is active
– The load has not been changed
within the last 2 minutes
– The scale has not been operated
within the last 2 minutes
– The load on the scale does
not exceed 2% of the maximum
capacity
* including the Signatories
of the Agreement on the European
Economic Area
42
When these requirements are met,
C is displayed in the measured
value line.
If the scale is not operated and the
load is not changed, internal
On
calibration and adjustment starts after
15 seconds have elapsed.
In the Setup menu, you can configure
the scale so that after calibration and
adjustment
– the application program
is restarted
On and reset
the application
– the application program remains
at its previous status
On without resetting
the app.
Also in Setup, you can configure
the scale so that it displays
a calibration prompt, but does not
perform the calibration/adjustment
functions automatically
Only adjustment prompt