Sartorius Factory Series FC06BBE-S Operating Instructions Manual page 110

Industrial weighing tech
Hide thumbs Also See for Factory Series FC06BBE-S:
Table of Contents

Advertisement

Display/Problem
Err 31
Print fct. blocked
Ref.wt. too light
Cannot update
Not a number
xxxxx
Too low
xxxxx
Too high
Too many char.
Err 10
x
:
1
x =
:
2
x =
:
x =
3
:
x =
4
"Checkerboard" pattern displayed
continuously
Err 340
No WP
blocked
The special code
J
remains displayed
The weight readout
changes constantly
The weight readout is
obviously wrong
If any other errors occur, contact your local Sartorius Service Center!
110
Cause
Interface handshake interrupted
(XOFF, CTS)
Error in storing reference weight
(with the counting or weighing-
in-percent application)
Reference updating not possible
(with the counting application)
Input wrong (with any
application program), e.g.,
alphabetic input not allowed
Input text too long
Key is stuck
Key pressed when switching on
the scale:
N (F1, F2, F5, F6), c
D, N (F3), 0, 3, 4, 9
2, 5, 6, ., p, t-right
1, 7, 8, N (F4), a,
t-left
M key was pressed when turning on
the scale, or is stuck
Operating parameter (EEPROM)
is wrong
Weighing platform is defective
Function blocked
None of the keys has been pressed
since the scale was turned on
Unstable ambient conditions
Too much vibration, or the scale
is exposed to a draft
A foreign object is caught between
the pan and the scale housing
The scale has not been
calibrated/adjusted
The scale was not tared
before weighing
The scale is not level
The dust cover is caught under the
weighing pan
Solution
Transmit XON, then CTS
Weight too light or there is
no sample on the scale
See "Counting" in "Operating
the Scale" for reference
updating criteria
Follow the instructions for the
application programs
Allowable text lengths, incl.
decimal point:
– S ID, NUM, L ID,
ID: max. 20 characters
– W ID: max. 14 characters
Release key
or
Contact your local
Sartorius Service Center
Contact your local Sartorius
Service Center
Contact your local Sartorius
Service Center
none
Press a key
Set up the scale
in another area
Change Setup configurations
to adapt the scale to the
ambient conditions
Remove the foreign object
Calibrate/adjust the scale
Tare before weighing
Level the scale
See "Replacing the Dust Cover" in
the chapter "Care and Maintenance"

Hide quick links:

Advertisement

Table of Contents
loading