Determination of the
Reproducibility (reproTEST)
Definition
Reproducibility is the ability of the
scale to display identical readouts
when it is loaded several times
with the same weight under constant
ambient conditions.
The standard deviation for a given
number of measurements is used
to quantify the reproducibility.
Check the Reproducibility of the Scale
Step
––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1. If
is set:
reproTEST
and proceed with step 4.
If
Selection mode
2. Select reproTEST
3. Start reproTEST
4. Number of measurements
is displayed; 6 measurements will
now be performed
The standard deviation
is displayed
5. End reproTEST
or restart reproTEST
Purpose
The "reproTEST" function auto-
matically determines the
reproducibility of results (based on
6 individual measurements).
In this way, the scale determines one
of the most important quantities
in relation to the place of installation.
The results are displayed with the
scale's accuracy.
Key(s) (or instruction)
soft key
isoTST
is set:
soft key
isoTST
soft key
Select
soft key
Start
soft key
End
soft key
Start
Operating the Scale
Preparation
§ Turn on the scale:
Press e/[ON/OFF]
> The Sartorius logo is displayed
> The scale performs a self-test
§ Select reproTEST in the Setup
menu: Press M
§ Select the Scale menu:
soft key
Menu
§ Select either
reproTEST
or
Selection mode
(factory setting): see "Configuring
the Scale"
§ Exit the Setup menu:
Press M
Display/Output
6
6
Max4200 g
0%
F
OO6
S
CAL: reproTEST
Start
d= 0.01g
100%
A
H
a
g
L
s
S
T
Wahl