Additional Functions
(in the Scale menu)
Password
You can block access to parameter
settings in the Setup menu and to
the ID-code input function, as well as
to the exact calibration weight, by
assigning a password.
Enter the password under Setup:
Input. See "Configuring the Scale"
for a detailed description.
Protecting Menu Parameters
In the Setup menu, you can define
whether menu parameters are
– accessible for changes
(
Change parameters,
)
[8-1-1]
– can be read only (
Read
parameters, [8-1-2]
Acoustic Signal
An acoustic signal is emitted when
you press a key. When the key
pressed is allowed, the signal
is a single beep-tone; when it is
not allowed, this is signaled by
a double-beep (key does not initiate
a function). In the Setup menu,
you can configure whether
– the acoustic signal should sound
(
,
)
On
[8-2-1]
– the acoustic signal should not
sound (
,
Off
[8-2-2]
Universal Switch for Remote Control
You can connect an external univer-
sal switch to the interface port of
your scale (e.g., a foot switch) for
remote control of the functions listed
below. In the Setup menu, you
can configure which function is to
be controlled via remote switch:
86
Function
p PRINT key
t key
q key
N F1 function key
c key
N F2 function key
See the "Pin Assignment Chart"
in this manual for further information.
Display Backlighting
You can have the display
backlighted for improved readability
of displayed values. In the Setup
menu, you can configure whether the
– display backlighting is on
(
,
On
[8-5-1]
– display backlighting shuts off
automatically after 4 minutes
without activity
)
(
Auto off after
4 minutes, [8-5-3]
Power-On Mode
You can configure the scale
so that when a power supply
is connected,
– the scale is off
(
Off/on/standby [8-6-1]
or
Off/on [8-6-2]
– the scale switches on
automatically (
)
[8-6-4]
You can also set the configurations
)
so that when the scale is turned off
after use, it is
– off (
Off/on
possible with scales that have a
weighing capacity ≥16 kg)
– in the standby mode (
,
standby
Menu item
After you turn on the scale,
a self-test of the functions is run
[8-4-1]
(
TEST
[8-4-2]
the bar graph is shown)
[8-4-3]
After the self-test has been
[8-4-4]
completed, the weighing range of
[8-4-5]
the scale is displayed (line for
[8-4-6]
metrological data shows different
increments)
Automatic Shutoff
When parameter
selected in the Setup menu, you can
configure whether
– the scale shuts off automatically
)
– automatic shutoff is deactivated
)
Printing an ISO/GLP/GMP-
compliant Record
In the Setup menu, you can configure
whether
– no ISO/GLP/GMP-compliant
– an ISO/GLP/GMP-compliant
)
,
Auto on
– every printout will be an ISO/
,
) (not
[8-6-2]
Undoing All Parameter Changes –
Reset Function
Off/on/
There is a factory setting for each
)
[8-6-1]
parameter. In the Setup menu, you
can configure whether
– factory settings will be restored
is displayed in the text line;
is
[8-6-2]
after 4 minutes without use
(
,
After 4 minutes
)
[8-7-1]
(
,
)
Off
[8-7-2]
record will be printed
(
,
)
Off
[8-10-1]
record will be printed only after
calibration/adjustment
(
Only for
calibration/
adjustment, [8-10-2]
GLP/GMP-compliant record
(
,
Always on
[8-10-3]
after exiting Setup (