External Calibration/Adjustment* with
a User-Defined Calibration Weight
First set either
Ext. cal./
adj.;user-defined wt.
(factory
Selection mode
setting) in the Setup: Scale menu.
You can define a weight for
calibration/adjustment. External
calibration/adjustment must
be performed with weights that are
traceable to a national standard
Define the Calibration Weight
Step
––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1. Select Setup
2. Select Input
3. Select calibration/adjustment
weight
4. Enter calibration weight
(e.g., 4000.00 g)
and store
6. Exit the Setup menu
* for verified scales, only "external adjustment" is possible
and that have error limits which are
or
at least 1/3 of the required
tolerance of the display accuracy.
The defined weight must equal
at least 10% of the maximum
scale capacity.
See page 40 for the external
calibration/adjustment sequence.
For this example, select external
calibration/adjustment with a user-
defined weight.
The scale has a factory-set weight
value (see "Specifications").
Key(s) (or instruction)
M
soft key
Input
q soft key 3 times
4 0 0 0 . 0
0
l soft key
M
Operating the Scale
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value
manually (see "Specifications")
Display/Output
3000.00 = last setting selected
41