TRIGger

TRIGger[:IMMediate]

Comments
Usage

TRIGger:SOURce

Parameters
Comments
Note
396 VT1422A Command Reference

TRIGger[:IMMediate]

causes one trigger when the module is set to the TRIG:SOUR
BUS or TRIG:SOUR HOLD mode.
This command is equivalent to the *TRG common command or the IEEE-488.2
"GET" bus command.
Related Commands: TRIG:SOURCE
Use VXIplug&play function: hpe1422_trigImm(...)
TRIG:IMM
TRIGger:SOURce <trig_source>
the trigger event.
Parameter
Parameter
Name
Type
trig_source
discrete (string)
The following table explains the possible choices.
BUS
EXTernal
HOLD
IMMediate
SCP
TIMer
TTLTrg<n>
The ARM system only exists while TRIG:SOUR is TIMer. When TRIG:SOUR
is not TIMer, SCPI compatibility requires that ARM:SOUR be IMM or an
Error -221,"Settings conflict" will be generated.
Use TRIGGER to start a measurement scan
configures the trigger system to respond to
Range of
BUS | EXT | HOLD | IMM | SCP | TIM
| TTLTrg<n>
TRIGger[:IMMediate], *TRG, GET (for GPIB)
"TRG" signal on terminal module
TRIGger[:IMMediate]
The trigger event is always satisfied.
SCP Trigger Bus (future SCP Breadboard)
The internal trigger timer
The VXIbus TTLTRG lines (n=0 through 7)
Default
Values
Chapter 7
Units
none