OUTPut:TYPE
Parameters
Comments
Usage
OUTPut:TYPE?
Parameters
Comments
Chapter 7
OUTPut:TYPE <select>,(@<ch_list>)
SCPs with programmable channels.
Parameter
Parameter
Name
Type
select
discrete (string)
ch_list
string
•
If the channels specified are on an SCP that doesn't support this function an error
will be generated. See the SCP's User's Manual to determine its capabilities.
•
PASSive configures the digital channel/bit to be passive (resistor) pull-up to
allow more than one output to wire-OR'd together.
•
ACTive configures the digital channel/bit to both source and sink current.
•
Related Commands: SOURce:PULSe:POLarity, OUTPut:TYPE?
•
*RST Condition: OUTP:TYPE ACTIVE (for TTL compatibility)
•
Send with VXIplug&play Function: hpe1422_cmd(...)
OUTP:TYPE PASS,(@140:143)
OUTPut:TYPE? <channel>
digital channel.
Parameter
Parameter
Name
Type
channel
string
•
The <channel> parameter must specify a single channel.
•
If the channel specified is not on a digital SCP, an error will be generated.
•
Returned Value: returns PASS or ACT. The type is string.
•
*RST Condition: returns ACT
•
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)
sets the output drive characteristic for digital
Range of
Values
PASSive | ACTive
100 - 163
make channels 40 to 43 passive pull-up
returns the output drive characteristic for a
Range of
Values
100 - 163