Usage

[SENSe:]FUNCtion:TEMPerature

Parameters
Comments
Chapter 7
Related Commands: CAL:REMote?,[SENSE:]STRAIN...
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set
to FBEN on all strain channels
Send with VXIplug&play Function: hpe1422_cmd(...)
SENS:FUNC:STRAIN:QUAR:POST 1,(@10000:10007) quarter bridge conversion for VT1529B
SENSE:FUNC:STRAIN:FBEN:POST 1,(@10800:10931) full bridge conversion for VT1529B
[SENSe:]FUNCtion:TEMPerature <type>,<sub_type>,[<range>,](@<ch_list>)
links channels to an temperature EU conversion based on the sensor specified in
<type> and <sub_type>. Note: Do not include the thermocouple reference
temperature channels in this command (for that, use the SENS:REF
<type>,<sub_type>,(@<channel>) command).
Parameter
Parameter
Name
Type
type
discrete (string)
sub_type
numeric (float32)
numeric (float32)
discrete (string)
range
numeric (float32)
ch_list
channel list (string)
Resistance temperature measurements (RTDs and THERmistors) require the use
of Current Source Signal Conditioning Plug-Ons. The following table shows the
Current Source setting that must be used for the following RTDs and Thermistors:
MAX (488 µA)
MIN (30 µA)
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and
16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified larger than one of the first four ranges,
the VT1422A selects the next higher range (for example, 4.1 selects the 16 V dc
range). Specifying a value larger than 16 generates an error. Specifying 0 selects
the lowest range (0.0625 V dc). Specifying AUTO selects auto range. The default
range (no range parameter specified) is auto range.
channels 10000 to 10007
channels 10800 to 10931
(64 channels)
Range of
Values
RTD | THERmistor | TCouple
for RTD, use 85 | 92
for THER, use 2250 | 5000 | 10000
for TC, use CUSTom | E | EEXT |
J | K | N | R | S | T
see comments
100 - 163
for RTD and THER,2250
for THER,5000 and THER,10000
VT1422A Command Reference