[SENSe:]FUNCtion:CUSTom:HVOLtage

Parameters
Comments
Usage
Chapter 7
[SENSe:]FUNCtion:CUSTom:HVOLtage [<range>,](@<ch_list>)
high-level input (DCV measurement) channels with the custom Engineering Unit
Conversion table loaded with the DIAG:CUST:MXB or DIAG:CUST:PIECE
commands.
This command is valid only on the VT1529B and is available with VXIplug&play
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).
Parameter
Parameter
Name
Type
range
numeric (float32)
ch_list
channel list (string)
See "Creating and Loading Custom EU Conversion Tables" on page 150.
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and
16 V dc. To select a range, simply specify the range value
(for example, 4 selects the 4 V dc range). If a value is specified that is larger
than one of the first four ranges, the VT1422A selects the next higher range (for
example, 4.1 selects the 16 V dc range). Specifying a value larger than 16 causes
an error -222 "Data out of range." Specifying 0 selects the lowest range
(0.0625 V dc). Specifying AUTO selects auto range. The default range (no range
parameter specified) is auto range.
When Accepted: Not while INITiated
Related Commands: DIAG:CUST:...
*RST Condition: all custom EU tables erased
Send with VXIplug&play Function: hpe1422_cmd(...)
DIAG:CUST:MXB slope,offset,(@10000)
SENS:FUNC:CUST:HVOL 1,(@10000)
INITiate then TRIGger module
Range of
Values
see first comment
10000 - 15731
send M*X+B to VT1422A for ch 10000
link custom EU with ch 10000
VT1422A Command Reference
[SENSe]
links VT1529B
Default
Units
V dc
none
341