Calibration:configure:voltage; Calibration:configure:voltage <Range>,<Zero_Fs - VXI VT1422A User Manual

Remote channel multi-function dac module user’s and scpi programming manual
Table of Contents

Advertisement

CALibration
Comments
Command
Sequence

CALibration:CONFigure:VOLTage

Parameters
Comments
264 VT1422A Command Reference
Related Commands: CAL:VAL:RES, CAL:STOR ADC
When Accepted: Not while INITiated
Send with VXIplug&play Function: hpe1422_cmd(...)
CAL:CONF:RES
*OPC? or SYST:ERR?
(now measure ref resistor with external DMM)
CAL:VAL:RES <measured value>
CAL:STORE ADC
CALibration:CONFigure:VOLTage <range>,<zero_fs>
voltage reference to the Calibration Bus. A measurement of the source voltage can
be made with an external multimeter connected to the H Cal and L Cal terminals on
the Terminal Module or the V H, V L, Ω H, and Ω L terminals on the Cal Bus
connector when not using a terminal module. The <range> parameter controls the
voltage level available when the <zero_fs> parameter is set to FSCale (full scale).
Parameter
Parameter
Name
Type
range
numeric (float32)
zero_fs
discrete (string)
The <range> parameter must be within ±5% of one of the five following values:
0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, 16 V dc. This parameter may be
specified in millivolts (mv).
The expected FSCALE output voltage of the calibration source will be
approximately 90% of the nominal value for each range, except the 16 V range
where the output is 10 V.
When Accepted: Not while INITiated
Related Commands: CAL:VAL:VOLT, STOR ADC
Send with VXIplug&play Function: hpe1422_cmd(...)
connect reference resistor to Calibration
Bus
must wait for CAL:CONF:RES to
complete
Send measured value to module
Store cal constants in non-volatile
memory (used only at end of complete
cal sequence)
connects the on-board
Range of
Values
see comments
ZERO | FSCale
Default
Units
volts
none