Parameters
Comments
Usage

OUTPut:CURRent:AMPLitude?

Parameters
Comments
Chapter 7
Parameter
Parameter
Name
Type
amplitude
numeric (float32)
ch_list
channel list (string)
Select 488E-6 (or MAX) for measuring resistances of less than 8000 Ω. Select
30E-6 (or MIN) for resistances of 8000 Ω and above. <amplitude> may be
specified in µA (ua).
For resistance temperature measurements ([SENSe:]FUNCtion:TEMPerature)
the Current Source SCP must be set as follows:
MAX (488 μA)
MIN (30 μA)
When *CAL? is executed, the current sources are calibrated on the range
selected at that time.
When Accepted: Not while INITiated
Related Commands: *CAL?, OUTP:CURR:AMPL?
*RST Condition: MIN
Send with VXIplug&play Function: hpe1422_cmd(...)
OUTP:CURR:AMPL 488ua,(@116:123)
OUTP:CURR:AMPL 30E-6,(@105)
OUTPut:CURRent:AMPLitude? (@<channel>)
Current Source SCP channel specified by channel.
Parameter
Parameter
Name
Type
channel
channel list (string)
The <channel> parameter must specify a single channel only.
If <channel> specifies an SCP which is not a Current Source, a +3007, "Invalid
signal conditioning plug-on" error is generated.
Range of
Values
MIN | 30E-6 | MAX | 488E-6
100 - 163
for RTD,85 | 92 and THER,2250
for THER,5000 | 10000
Set Current Source SCP at channels 16
through 23 to 488 µA
Set Current Source SCP at channel 5 to
30 µA
returns the range setting of the
Range of
Values
100 - 163
VT1422A Command Reference
OUTPut
Default
Units
A dc
none
Default
Units
none