OUTPut:SHUNt?
Parameters
Comments
Usage
OUTPut:SHUNt:SOURce
Parameters
Comments
Chapter 7
OUTPut:SHUNt? (@<channel>)
specified Strain SCP or RSC channel.
Parameter
Parameter
Name
Type
channel
channel list (string)
•
The <channel> parameter must specify a single channel only.
•
If <channel> specifies a non strain SCP or RSC, a 3007 "Invalid signal
conditioning plug-on" error is generated.
•
Returned Value: Returns 1 or 0. The data type is uint16.
•
Related Commands: OUTP:SHUNT
•
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
OUTPUT:SHUNt? (@116)
OUTPUT:SHUNt? (@10124)
OUTPut:SHUNt:SOURce <select>,(@<ch_list>)
shunt resistance for a VT1529A/B Remote Strain Bridge Conditioning unit. The
VT1529A/B has an internal shunt resistor and also supports an external user supplied
resistor.
Parameter
Parameter
Name
Type
select
discrete (string)
ch_list
channel list (string)
•
If <ch_list> specifies a non VT1529A/B strain SCP, a 3007 "Invalid signal
conditioning plug-on" error is generated.
•
Only one channel on each VT1529A/B needs to be specified since a single
resistor is used for all channels in the module.
•
When Accepted: Not while INITiated
returns the status of the shunt resistance on the
Range of
Values
100 - 15731
Check status of shunt resistance on
channel 16
Check status of shunt resistance on
VT1529A/B channel 0124
selects the source of the bridge
Range of
Values
INTernal | EXTernal
10000 - 15731
VT1422A Command Reference
OUTPut
Default
Units
none
Default
Units
none
none
317