CALibration:TARE?

Command
Sequence

CALibration:VALue:RESistance

Parameters
Comments
Chapter 7

CALibration:TARE?

Returns a value to indicate the success of the last CAL:TARE
operation. CAL:TARE? returns the value only after the CAL:TARE operation
is complete.
Returned Value:
Value
0
-1
-2
The data type for this returned value is int16.
Executing CAL:TARE sets the Calibrating bit (bit 0) in Operation Status Group.
Executing CAL:TARE? resets the bit.
Related Commands: CAL:STOR TARE
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
CAL:TARE <ch_list>
CAL:TARE?
CAL:STORE TARE
CALibration:VALue:RESistance <ref_ohms>
the on-board reference resistor to the module for A/D calibration.
Parameter
Parameter
Name
Type
ref_ohms
numeric (float32)
Use the CAL:CONF:RES command to configure the reference resistor for
measurement at the Calibration Bus connector.
A four-wire measurement of the resistor can be made with an external multimeter
connected to the HCAL, LCAL, HOHM and LOHM terminals on the Terminal
Module or the V H, V L, Ω H, and Ω L terminals on the Cal Bus connector
when not using a terminal module.
Meaning
Cal OK
Cal Error
Query the Error Queue (SYST:ERR?)
See "Error Messages" on page 453.
No results available
to correct channel offsets
to return the success flag from the
CAL:TARE operation
Optional depending on necessity of long
term storage
sends the just-measured value of
Range of
7,500 ± 5%
VT1422A Command Reference
CALibration
Further Action
None
Also run *TST?
Perform CAL:TARE
before CAL:TARE?
Default
Value
Units
ohms