[Sense:]Function:hvoltage; [Sense:]Function:hvoltage [<Range>,](@<Ch_List>) - VXI VT1422A User Manual

Remote channel multi-function dac module user’s and scpi programming manual
Table of Contents

Advertisement

[SENSe:]FUNCtion:HVOLtage

Parameters
Comments
Usage
Chapter 7
[SENSe:]FUNCtion:HVOLtage [<range>,](@<ch_list>)
VT1529B channels to return dc voltage on the high-level input (Excitation Sense +
and -) pins on the RJ-45 connector for the channel.
This command is valid only on the VT1529B and is available with VXIplug&play
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).
Parameter
Parameter
Name
Type
range
numeric (float32)
ch_list
channel list (string)
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and
16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified that is larger than one of the first four
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the
16 V dc range). Specifying a value larger than 16 causes an error. Specifying 0
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range.
The default range (no range parameter specified) is auto range.
The VT1529A/B has a fixed gain of 32. Keep this in mind when <range> is set.
See "DCV Measurement Command Sequence" on page 177.
When Accepted: Not while INITiated
Related Commands: CAL:REMote?
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set
to FBEN on all strain channels.
Send with VXIplug&play Function: hpe1422_cmd(...)
FUNC:HVOL (@10000:10001)
links the specified
Range of
Values
see comments
10000 - 15731
Channels 0 - 1 measure high-level input
voltage in auto-range (defaulted)
VT1422A Command Reference
[SENSe]
Default
Units
V dc
none
345